A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Project Subscriptions

Vendors Products
Smiths-medical Subscribe
Medfusion 4000 Wireless Syringe Infusion Pump Subscribe
Advisories
Source ID Title
EUVD EUVD EUVD-2017-4264 A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Fixes

Solution

No solution given by the vendor.


Workaround

No workaround given by the vendor.

History

No history.

Projects

Sign in to view the affected projects.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2024-08-05T18:43:56.615Z

Reserved: 2017-08-09T00:00:00.000Z

Link: CVE-2017-12725

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2018-02-15T10:29:00.523

Modified: 2024-11-21T03:10:06.197

Link: CVE-2017-12725

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses